A liquid chromatographic method for the
determination of fumonisins B1 (FB1) and B2
(FB2) in corn-based foods for infants and young
children was subjected to an interlaboratory
validation study involving 11 laboratories. Five
blind duplicate sample pairs of each matrix
were analyzed to establish the accuracy,
repeatability, and reproducibility of the method.
Mass fractions in the baby food samples
ranged from 89.1 to 384.4 ?g/kg FB1 and from
22.5 to 73.6 ?g/kg FB2. The method involved a
warm extraction with citrate phosphate buffer-
methanol-acetonitrile (50 + 25 + 25, v/v/v), a
cleanup through an immunoaffinity column,
and an end-determination of fumonisins by LC
after automated precolumn derivatization with
o-phthaldialdehyde reagent. RSDs for withinlaboratory
repeatability (RSDr) ranged from 6.8
to 23.5% for FB1 and 7.6 to 22.9% for FB2. RSDs
for between-laboratory reproducibility (RSDR)
ranged from 15.4 to 26.2% for FB1 and 21.6 to
36.3% for FB2. Mean FB1 recoveries from baby
foods spiked at 100.0 and 250.0 ?g/kg were 89
and 96%, respectively; for FB2 spiked foods at
25.0 and 62.5 ?g/kg recoveries were 90 and 85%,
respectively. HorRat values ranged from 0.8 to
1.2 for FB1, whereas for FB2 they ranged from
0.9 to 1.4 when calculated according to Horwitz,
and from 1.0 to 1.7 when calculated according
to Thompson, indicating an acceptable amonglaboratory
precision for all matrixes (HorRat
values <2).