Up today Hamamatsu R3292 is the Position Sensitive Photo
Multiplier Tube (PSPMT) with the largest sensitive area (10 cm of
diameter). At the same time it has the minimum size for clinical
application in Nuclear Medicine. A portable gamma camera was realized,
based on 5" PSPMT coupled to a scintillating array. The head has a light
weight (15 kg.) spatial resolution resulted better than that of Anger
Camera with good linearity response, good energy resolution and FOV
coincident with intrinsic one of PSPMT. To optimize gamma camera
response two different scintillating arrays were tested: YAP:Ce and CsI
(Tl). Their overall size cover all photocathode active area, and crystal
pixel size was 2 mm×2 mm. The detection efficiency resulted
comparable to that of Anger Camera. The best result was obtained by CsI
(Tl) scintillating: an intrinsic spatial resolution of 1.6 mm FWHM and a
relative energy resolution of 17% FWHM. With a standard general purpose
collimator a spatial resolution of about 2 mm resulted. Some preliminary
results were also obtained in breast scintigraphy