Data di Pubblicazione:
2014
Abstract:
Two different approaches were tested in this study to suppress sulfide interference in the
silicomolybdate method of silica determination, namely: oxidation of sulfide to elemental sulfur; and conversion of the sulfide species to H2S and stripping by air bubbling. Based on the results obtained the latter approach is deemed to represent the better and quicker option to eliminate sulfide interference in silica colorimetric analysis.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
sulfide interference; colorimetric analysis; silica
Elenco autori:
Lelli, Matteo
Link alla scheda completa:
Pubblicato in: