Skip to Main Content (Press Enter)

Logo CNR
  • ×
  • Home
  • Persone
  • Pubblicazioni
  • Strutture
  • Competenze

UNI-FIND
Logo CNR

|

UNI-FIND

cnr.it
  • ×
  • Home
  • Persone
  • Pubblicazioni
  • Strutture
  • Competenze
  1. Pubblicazioni

A graphene-based pH sensor on paper for human plasma and seawater

Contributo in Atti di convegno
Data di Pubblicazione:
2019
Abstract:
The relevance of pH assessment in clinical analysis, environmental and industrial control, has raised the demand for the development of portable, low cost and easy-to-use monitoring systems. This paper proposes a pH sensor printed on a paper support passivated with a solid-ink coating. The sensor exploits the pH sensitivity of a reduced graphene oxide functionalized with 3-(4-aminophenil)propionic acid. The sensor responded in the pH range [4, 10] and had a sensitivity of 46 mV/pH. Tests on human plasma and seawater proved this pH sensor to have similar performances than those of a commercial pH-meter with an uncertainty of 0.1 and 0.2 pH unit in plasma and seawater, respectively.
Tipologia CRIS:
04.01 Contributo in Atti di convegno
Keywords:
biosensor; seawater; human plasma; graphene; pH
Elenco autori:
Salvo, Pietro
Autori di Ateneo:
SALVO PIETRO
Link alla scheda completa:
https://iris.cnr.it/handle/20.500.14243/370530
Pubblicato in:
IEEE ENGINEERING IN MEDICINE AND BIOLOGY ... ANNUAL CONFERENCE PROCEEDINGS
Series
  • Dati Generali

Dati Generali

URL

http://www.scopus.com/record/display.url?eid=2-s2.0-85077893745&origin=inward
  • Utilizzo dei cookie