Performance of a time-of-flight silicon strip telescope in the temperature range 20°C to -55°C
Academic Article
Publication Date:
1995
abstract:
A cold silicon strip telescope for time-of-flight determination at -55°C has been tested using a hadron beam at Saturne II. Saclay. We present performance tests of the telescope, the time-of-flight distributions between pairs of silicon hodoscopes and their dependence on temperature. A detailed description of the apparatus, the refrigerator system and the calibration procedures is also reported. A linear relationships between the time resolution and the temperature of the apparatus is measured. An improvement of a factor 2 in the time-of-flight resolution is observed when the temperature decreases from +20°C to -55°C. © 1995.
Iris type:
01.01 Articolo in rivista
List of contributors: