Development of stable and reproducible biosensors based on electrochemical impedance spectroscopy: Three-electrode versus two-electrode setup
Articolo
Data di Pubblicazione:
2014
Abstract:
This work focuses on the development of electrochemical impedance biosensors based on capacitance readout, for the detection of biomolecules in small sample volumes. We performed electrochemical impedance spectroscopy (EIS) measurements of DNA hybridization in electrochemical cells with microfabricated gold electrodes. The time stability of the device was tested in two different configurations: two microelectrodes in a microfluidic channel; two microelectrodes plus a reference electrode in an electrochemical cell. Our results demonstrate that the three-electrode setup is more stable, more reproducible, and suitable for real-time measurements. In the last part of the work we perform a test study of DNA hybridization in real time, and we show that the three-electrode configuration can measure the process in situ and in real time. (C) 2013 Elsevier B.V. All rights reserved.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
DNA self-assembling; DNA hybridization; EIS; Real-time biosensors
Elenco autori:
Grenci, Gianluca; Tormen, Massimo
Link alla scheda completa:
Pubblicato in: