Temperature dependence of dc SQUID magnetometer performances

Articolo
Data di Pubblicazione:
2004
Abstract:
We report experimental results on temperature dependence of the main characteristics of fully integrated DC SQUID magnetometers realized on niobium technology. At T ΒΌ 4:2 K the sensor shown a white magnetic field noise spectral density of 2.9 fT/Hz 1/2 . A slow increase of the field noise (about 20%) was observed with increasing temperature up to 5 K, giving a considerable tolerance of the working temperature of niobium magnetometers in some innovative multichannel systems for magnetoencephalography.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Magnetometer; Superconducting device; SQUID
Elenco autori:
Granata, Carmine; Lisitskiy, Mikhail; Russo, Maurizio
Autori di Ateneo:
GRANATA CARMINE
LISITSKIY MIKHAIL
Link alla scheda completa:
https://iris.cnr.it/handle/20.500.14243/118196
Pubblicato in:
JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS