Data di Pubblicazione:
2006
Abstract:
Usefulness of formulas to predict effective CPAP level for OSAS treatment is under
debate.Inthisstudy,CPAPlevelscalculatedaccordingtofourdifferentformulas
[StradlingJRetal,RespirMed(2004)98:152-4;HoheiselGBetal,AmJRespir
CritCareMed(1994)149:A496;SérièsFetal,AmJRespirCritCareMed(2000)
162:94-7;HoffsteinVetal,AmJRespirCritCareMed(1994)150:486-8]were
comparedtoeffectiveCPAPleveltitratedduringpolysomnography.CPAPtitration
waspreformedin136OSASsubjects(22F,age54 ± 11 yrs,BMI34.6 ± 6.1
kg/m2 , AHI58 ± 22) bymeansofoneAPAPdevice(AutosetT,ResMed)applied
duringapolysomnographicstudy.Aseffectivepressurewasconsideredthe95th
centile pressureleveldeliveredbytheAPAP,correctedaccordingtotheanalysisof
polysomnographicsignals.Correlationsbetween calculatedandeffectivepressures
werealwayssignificant,butloose(r2 between 0.044and0.139).Differences <=2
cmH2O fromeffectivepressurewerefoundfor78%pressurevaluescalculated
by Stradling'sformula,andfor19-46%valuescalculatedbytheotherformulas.
Differencesbetween95th centilepressuresandcalculatedpressureswereslightly
higher (64%pressurelevelsbyStradling'sformula,and14-32%bytheotherones,
differing <=2 cmH2O from95th centilepressures).Differencesbetweeneffective
pressureandpressurecalculatedbyStradling'sformularangedbetween-3.5and
+6.5 cmH2O. Onlyvaluescalculatedbyoneofthetestedpredictiveformulas
approximatedsufficientlytotherealeffectivepressureinahighpercentageof
cases, butevenbythatformulaeffectivepressureisnotestimatedwithacceptable
accuracyin1outof5cases.