Data di Pubblicazione:
2015
Abstract:
A miniaturized optical biosensor for low-level fluorescence emitted by DNA strands labelled with CY5 is showed. Aim of this work is to demonstrate that a Si-based photodetector, having a low noise and a high sensitivity, can replace traditional detection systems in DNA-microarray applications. The photodetector used is a photomultiplier (SiPM), with 25. pixels. It exhibits a higher sensitivity than commercial optical readers and we experimentally found a detection limit for spotted dried samples of ~1 nM. We measured the fluorescence signal in different operating conditions (angle of analysis, fluorophores concentrations, solution volumes and support). Once fixed the angle of analysis, for samples spotted on Al-TEOS slide dried, the system is proportional to the concentration of the analyte in the sample and is linear in the range 1. nM-1. ?M. For solutions, the range of linearity ranges from 100. fM to 10. nM. The system potentialities and the device low costs suggest it as basic component for the design and fabrication of a cheap, easy and portable optical system.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
DNA microarray; Fluorophore detection; Optical Biosensor; SiPM
Elenco autori:
Santangelo, MARIA FRANCESCA; Sciuto, EMANUELE LUIGI; Libertino, Sebania
Link alla scheda completa:
Pubblicato in: