Skip to Main Content (Press Enter)

Logo CNR
  • ×
  • Home
  • Persone
  • Pubblicazioni
  • Strutture
  • Competenze

UNI-FIND
Logo CNR

|

UNI-FIND

cnr.it
  • ×
  • Home
  • Persone
  • Pubblicazioni
  • Strutture
  • Competenze
  1. Pubblicazioni

High throughput detection chain for time domain optical mammography

Articolo
Data di Pubblicazione:
2018
Abstract:
A novel detection chain, based on 8 Silicon Photomultipliers (forming a wide-area custom-made detection probe) and on a time-to-digital converter, was developed to improve the signal level in multi-wavelength (635-1060 nm) time domain optical mammography. The performances of individual components and of the overall chain were assessed using established protocols (BIP and MEDPHOT). The photon detection efficiency was improved by up to 3 orders of magnitude, and the maximum count rate level was increased by a factor of 10 when compared to the previous system, based on photomultiplier tubes and conventional time-correlated single-photon counting boards. In the estimate of optical parameters, the novel detection chain provides performances comparable to the previous system, widely validated in clinics, but with higher signal level, higher robustness, and at a lower price per channel, thus targeting important requirements for clinical applications. (C) 2018 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
diffuse optical tomography
Elenco autori:
Pifferi, ANTONIO GIOVANNI; Taroni, Paola
Link alla scheda completa:
https://iris.cnr.it/handle/20.500.14243/401410
Pubblicato in:
BIOMEDICAL OPTICS EXPRESS
Journal
  • Dati Generali

Dati Generali

URL

https://www.osapublishing.org/boe/fulltext.cfm?uri=boe-9-2-755&id=381045
  • Utilizzo dei cookie

Realizzato con VIVO | Designed by Cineca | 26.5.0.0 | Sorgente dati: PREPROD (Ribaltamento disabilitato)