Data di Pubblicazione:
2015
Abstract:
Fast neutron measurements were performed on the VESUVIO beam-line at the ISIS spallation source using a new telescope proton recoil spectrometer. Neutrons interact on a plastic target. Proton production is mainly due to elastic scattering on hydrogen nuclei and secondly due to interaction with carbon nuclei. Recoil protons are measured by a proton spectrometer, which uses in coincidence a 2.54 cm thick YAP scintillator and a 500 mu m thick silicon detector, measuring the full proton recoil energy and the partial deposited energy in transmission, respectively. Recoil proton spectroscopy measurements (up to E-p = 60 MeV) have been interpreted by using Monte Carlo simulations of the beam-line. This instrument is of particular interest for the characterization of the ChipIr beam-line at ISIS, which was designed to feature an atmospheric-like neutron spectrum for the irradiation of micro-electronics.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
SINGLE-EVENT UPSET; DIAMOND DETECTOR; CROSS-SECTIONS; SCINTILLATORS; FUSION; SPECTROSCOPY
Elenco autori:
Gorini, Giuseppe; Nocente, Massimo; Croci, Gabriele; Cazzaniga, Carlo; Rebai, Marica; Tardocchi, Marco
Link alla scheda completa:
Pubblicato in: