Torsional and Cyclic Fatigue Resistance of a New Nickel-Titanium Instrument Manufactured by Electrical Discharge Machining

Articolo
Data di Pubblicazione:
2016
Abstract:
The purpose of this study was to evaluate the torsional and cyclic fatigue resistance of the new Hyflex EDM OneFile (Coltene/Whaledent AG, Altstatten, Switzerland) manufactured by electrical discharge machining and compare the findings with the ones of Reciproc R25 (VDW, Munich, Germany) and WaveOne Primary (Dentsply Maillefer, Ballaigues, Switzerland). Methods: One hundred-twenty new Hyflex EDM One- File (#25/0.08), Reciproc R25, and WaveOne Primary files were used. Torque and angle of rotation at failure of new instruments (n = 20) were measured according to ISO 3630-1 for each brand. Cyclic fatigue resistance was tested measuring the number of cycles to failure in an artificial stainless steel canal with a 60 angle and a 3-mm radius of curvature. Data were analyzed using the analysis of variance test and the Student- Newman-Keuls test for multiple comparisons. The fracture surface of each fragment was examined with a scanning electron microscope. Results: The cyclic fatigue of Hyflex EDM was significantly higher than the one of Reciproc R25 and WaveOne Primary (P < .05 and P < .001, respectively). Hyflex EDM showed a lower maximum torque load (P < .05) but a significantly higher angular rotation (P < .0001) to fracture than Reciproc R25 and WaveOne Primary. No significant difference was found comparing the maximum torque load, angular rotation, and cyclic fatigue of Reciproc R25 and WaveOne Primary (P > .05). Conclusions: The new Hyflex EDM instruments (controlled memory wire) have higher cyclic fatigue resistance and angle of rotation to fracture but lower torque to failure than Reciproc R25 and WaveOne Primary files (M-wire for both files).
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Cyclic fatigue; electrical discharge machining; Hyflex EDM; M-wire; torsional resistance
Elenco autori:
Boninelli, SIMONA MARIA CRISTINA
Autori di Ateneo:
BONINELLI SIMONA MARIA CRISTINA
Link alla scheda completa:
https://iris.cnr.it/handle/20.500.14243/402469
Pubblicato in:
JOURNAL OF ENDODONTICS
Journal
  • Dati Generali

Dati Generali

URL

https://pubmed.ncbi.nlm.nih.gov/26586518/
  • Utilizzo dei cookie

Realizzato con VIVO | Designed by Cineca | 26.5.0.0 | Sorgente dati: PREPROD (Ribaltamento disabilitato)