Detection and compensation of organ/lesion motion using 4D-PET/CT respiratory gated acquisition techniques.
Articolo
Data di Pubblicazione:
2010
Abstract:
PURPOSE: To describe the degradation effects produced by respiratory organ and lesion motion on PET/CT images and to define the role of respiratory gated (RG) 4D-PET/CT techniques to compensate for such effects.
METHODS: Based on the literature and on our own experience, technical recommendations and clinical indications for the use of RG 4D PET/CT have been outlined.
RESULTS: RG 4D-PET/CT techniques require a state of the art PET/CT scanner, a respiratory monitoring system and dedicated acquisition and processing protocols. Patient training is particularly important to obtain a regular breathing pattern. An adequate number of phases has to be selected to balance motion compensation and statistical noise. RG 4D PET/CT motion free images may be clinically useful for tumour tissue characterization, monitoring patient treatment and target definition in radiation therapy planning.
CONCLUSIONS: RG 4D PET/CT is a valuable tool to improve image quality and quantitative accuracy and to assess and measure organ and lesion motion for radiotherapy planning.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Respiratory organ motion; Respiratory gated 4D-PET/CT; Radiotherapy