Millimeterwave reflectivity tests on MetOp-SG MWI On Board Calibration Target

Articolo
Data di Pubblicazione:
2021
Abstract:
he On-Board Calibration Target (OBCT) for the Micro-Wave Imager instrument on the MetOp-SG-B satellite is a microwave blackbody, operating between 14 and 220 GHz (extended frequency range), providing the MWI receivers with a warm calibration reference. It is designed for an emissivity better than 0,9999 over the MWI 26 channels (including dual polarization channels). The OBCT radiometric performance has been extensively verified in ISTP (formerly IFP) microwave laboratory, by measuring reflectivity over the extended bandwidth. Measurement techniques and results are described in this paper, showing that the OBCT models tested so far are fully compliant with the MWI requirements. © 2021 IOP Publishing Ltd and Sissa Medialab.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Microwave Calibrators; Microwave radiometers; Space instrumentation
Elenco autori:
Simonetto, Alessandro
Autori di Ateneo:
SIMONETTO ALESSANDRO
Link alla scheda completa:
https://iris.cnr.it/handle/20.500.14243/395041
Pubblicato in:
JOURNAL OF INSTRUMENTATION
Journal
  • Dati Generali

Dati Generali

URL

https://iopscience.iop.org/article/10.1088/1748-0221/16/03/P03036/meta