Data di Pubblicazione:
2001
Abstract:
Click-evoked otoacoustic emissions (CEOAEs) of full-term babies were considered. By means of the wavelet transform, each CEOAE was decomposed into frequency bands. The rms amplitude and test-retest correlation were computed from these bands. The results indicated that both the amplitude and correlation were not invariant with time but reached a maximum in specific time windows, depending on the frequency of the component. For all components, the correlation was greatly decreased for latencies >12.5 ms. As a result, comparison between the performance of the default ILO 88 window (2.5-20 ms) and the 2.5-12.5 ms window showed that for all frequencies in the 1.5-6 kHz range there was a statistically significant improvement in the correlation.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Time windowing Wavelet transform; Click-evoked otoacoustic emissions; Full-term neonates
Elenco autori:
Ravazzani, PAOLO GIUSEPPE; Grandori, Ferdinando; Tognola, Gabriella
Link alla scheda completa:
Pubblicato in: