Data di Pubblicazione:
2005
Abstract:
This work presents the results obtained with a new generation portable gamma camera, assembled with an innovative technique based on CsI(Tl) scintillation crystals inserted into the square holes of a tungsten collimator. Characterization measurements have been performed on three detectors with different crystal size, in order to obtain 7×7, 8×8 and 9×9 arrays over the fixed area of 25.7×25.7 mm2, having the same overall dimensions of a Position Sensitive Photomultiplier Tube. Detector sensitivity, energy resolution and spatial resolution have been measured, and the acquired image quality has been evaluated with particular attention to pixel identification capability.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
nuclear medicine imaging; scintillation detector; position sensitive photomultiplier tube
Elenco autori:
Massari, Roberto; Soluri, Alessandro
Link alla scheda completa: