Lymphocyte subset counting by immunoperoxidase using an automated routine hematologic analyzer
Articolo
Data di Pubblicazione:
1989
Abstract:
Using monoclonal antibodies against CD2, CD4, CD8 and CD19 antigens and an automated biotin-avidin immunoperoxidase technique on whole blood samples, we evaluated the technical performance and clinical usefulness of lymphocyte subset counting by the routine hematology analyzer Technicon H*1. Statistical evaluation demonstrated excellent precision and very good correlation with the immunofluorimetric flow cytometer Ortho Spectrum III. Correlation between manual immunofluorescence at the microscope and the H*1 method was much poorer, owing to the high intrinsic imprecision of the manual method. Reference ranges obtained with the H*1 immunoperoxidase method in 44 healthy subjects closely matched those obtained with the Spectrum III. In 46 subjects with or at risk for HIV infection, we found with the H*1 method a significant decrease in CD4+ cells and in the CD4+/CD8+ cell ratio, which was progressively more marked in HIV- negative patients with lymphadenopathic syndrome, AIDS-related complex, and in patients with full-blown AIDS.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Antibodies; Monoclonal/diagnostic use Antigens; Differentiation/analysis Flow Cytometry Fluorescent Antibody Technique HIV Seropositivity/pathology Humans Immunoenzyme Techniques/*instrumentation Leukocyte Count/*instrumentation Lymphocytes/*classification/enzymology Predictive Value of Tests
Elenco autori:
Quarantelli, Mario
Link alla scheda completa:
Pubblicato in: