Aim of this study was to perform a detailed clinical validation of a novel fully automatic method for vertebral
morphometry. About 80 spine lateral radiographs were evaluated both automatically, by the proposed algorithm, and
manually, by an experienced radiologist. The following metrics were used for algorithm performance assessment:
sensitivity and specificity in vertebra detection; errors in the localisation of characteristic points of vertebral border;
errors in the measurement of six diagnostic parameters; level of agreement and correlation between manual and
automatic morphometric measurements; overall accuracy of automatic diagnoses with respect to manual ones.
Obtained results showed a very good performance in vertebra detection (sensitivity = 89.1% and specificity = 100.0%).
Average errors in the localisation of vertebral characteristic points were always smaller than 3 mm (range 0.85-2.79mm),
causing relative errors in diagnostic parameter values ranging from -5.01 to +6.10%. Bland-Altman analysis documented
a mean error in automatic measurements of diagnostic ratios of 0.01 ± 0.18 (bias ± 2 SDs), while Pearson's correlation
coefficient resulted r = 0.71 (p < 0.001). Finally, an optimal diagnostic coincidence (92.8%) was found between automatic
and manual diagnoses. Therefore, the adopted method has a potential for an effective employment in clinical routine
for reliable diagnosis of vertebral fractures.