Forward planned intensity modulated radiotherapy (IMRT) for whole breast postoperative radiotherapy. Is it useful? When?

Articolo
Data di Pubblicazione:
2011
Abstract:
The purpose was to compare the dosimetric results observed in 201 breast cancer patients submitted to tangential forward intensity-modulated radiation therapy (IMRT) with those observed in 131 patients treated with a standard wedged 3D technique for postoperative treatment of whole breast, according to breast size and supraclavicular node irradiation. Following dosimetric parameters were used for the comparison: D(max), D(min), D(mean), V(95%) and V(107%) for the irradiated volume; D(max), D(mean), V(80%) and V(95%) for the ipsilateral lung; D(max), D(mean), V(80%) and V(95%) for the heart. Stratification was made according to breast size and supraclavicular (SCV) nodal irradiation. As respect to irradiated volume, a significant reduction of V(107%) (mean values: 7.0 ± 6.6 versus 2.4 ± 3.7, p < 0.001) and D(max) (mean % values: 111.2 ± 2.7 versus 107.7 ± 6.3, p < 0.001), and an increase of D(min) (mean % values: 65.0 ± 17.4 versus 74.9 ± 12.9, p < 0.001) were observed with forward IMRT. The homogeneity of dose distribution to target volume significantly improved with forward IMRT in all patient groups, irrespective of breast size or supraclavicular nodal irradiation. When patients treated with supraclavicular nodal irradiation were excluded from the analysis, forward IMRT slightly reduced V(80%) (mean values: 3.7 ± 2.6 versus 3.0 ± 2.4, p = 0.03) and V(95%) (mean values 1.9 ± 1.8 versus 1.2%± 1.5; p = 0.001) of the ipsilateral lung. The dose to the heart tended to be lower with IMRT but this difference was not statistically significant. Tangential forward IMRT in postoperative treatment of whole breast improved dosimetric parameters in terms of homogeneity of dose distribution to the target in a large sample of patients, independent of breast size or supraclavicular nodal irradiation. Lung irradiation was slightly reduced in patients not undergoing to supraclavicular irradiation.
Tipologia CRIS:
01.01 Articolo in rivista
Elenco autori:
DE GAETANO, Andrea; Panunzi, Simona
Autori di Ateneo:
DE GAETANO ANDREA
PANUNZI SIMONA
Link alla scheda completa:
https://iris.cnr.it/handle/20.500.14243/292820
Pubblicato in:
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS