Vibration maps of capacitive micromachined ultrasonic transducers by laser interferometry

Articolo
Data di Pubblicazione:
2002
Abstract:
A 1.8-mm /spl times/ 1.8-mm capacitive micromachined ultrasonic transducer (CMUT) element is experimentally characterized by means of optical measurements. Optical displacement measurements provide information on the resonant behavior of the single membranes and also allow us to investigate the dispersion in the frequency spectrum of adjacent membranes. In addition, higher order mode shapes are observed, showing that either symmetrical or asymmetrical modes are excited in CMUT membranes. Laser interferometry vibration maps, combined with quantitative displacement measurements, provide information about the quality and repeatability of the fabrication process, which is a basic requirement for 2D array fabrication for ultrasound imaging.
Tipologia CRIS:
01.01 Articolo in rivista
Elenco autori:
Foglietti, Vittorio
Autori di Ateneo:
FOGLIETTI VITTORIO
Link alla scheda completa:
https://iris.cnr.it/handle/20.500.14243/178295
Pubblicato in:
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL