Radiometric and Signal-to-Noise Ratio Properties of Multiplex Dispersive Spectrometry

Articolo
Data di Pubblicazione:
2010
Abstract:
Recent theoretical investigations have shown important radiometric disadvantages of interferential multiplexing in Fourier transform spectrometry that apparently can be applied even to coded aperture spectrometers. We have reexamined the methods of noninterferential multiplexing in order to assess their signal-to-noise ratio (SNR) performance, relying on a theoretical modeling of the multiplexed signals.We are able to show that quite similar SNR and radiometric disadvantages affect multiplex dispersive spectrometry. The effect of noise on spectral estimations is discussed.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Sagnac interferometer; Multiplex dispersive spectrometry; Fourier transform spectrometry; signal to noise ratio
Elenco autori:
Lastri, Cinzia; Marcoionni, Paolo; Nardino, Vanni; Guzzi, Donatella; Pippi, Ivan
Autori di Ateneo:
GUZZI DONATELLA
LASTRI CINZIA
NARDINO VANNI
Link alla scheda completa:
https://iris.cnr.it/handle/20.500.14243/42333
Pubblicato in:
APPLIED OPTICS