Energy resolution of gamma-ray spectroscopy of JET plasmas with a LaBr(3) scintillator detector and digital data acquisition
Articolo
Data di Pubblicazione:
2010
Abstract:
A new high efficiency, high resolution, fast gamma -ray spectrometer was recently installed at the JET tokamak. The spectrometer is based on a LaBr 3 (Ce) scintillator coupled to a photomultiplier tube. A digital data acquisition system is used to allow spectrometry with event rates in excess of 1 MHz expected in future JET DT plasmas. However, at the lower rates typical of present day experiments, digitization can degrade the energy resolution of the system, depending on the algorithms used for extracting pulse height information from the digitized pulses. In this paper, the digital and analog spectrometry methods were compared for different experimental conditions. An algorithm based on pulse shape fitting was developed, providing energy resolution equivalent to the traditional analog spectrometry method.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
neutron spectroscopy; gamma ray spectroscopy; nuclear fusion
Elenco autori:
Gorini, Giuseppe; Nocente, Massimo; Grosso, GIOVANNI MARIA; Murari, Andrea; Tardocchi, Marco; Pietropaolo, Antonino
Link alla scheda completa:
Pubblicato in: