  1. Pubblicazioni

Conditions for calibrConditions for calibration of an isothermal titration calorimeter using chemical reactions

Articolo
Data di Pubblicazione:
2013
Abstract:
The reaction of protonation of 2-amino-2-(hydroxymethyl)-1,3-propanediol (TRIS) is a suitable one for the calibration of isothermal titration calorimeter (ITC), providing that experimental conditions are appropriately chosen. The conditions and methods for handling experimental data from a nanowatt-ITC are discussed. Also, the binding of Ba2+ to 18-Crown-6 is successfully used to check the accuracy and precision of the chemical calibration performed with TRIS. This latter reaction has the additional advantage that the data can also be used for a check on the determination of the value of a binding constant. The anomaly of the first injection in ITC is analyzed and, by combining calorimetric and spectroscopic measurements, it is shown that it mainly results from a backlash effect of the syringe plunger rather than from a diffusion effect.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Backlash vs. diffusion; Calorimetry; Chemical calibration; ITC; Solution equilibria
Elenco autori:
Zito, VALERIA MARIA VINCENZA
Autori di Ateneo:
ZITO VALERIA MARIA VINCENZA
Link alla scheda completa:
https://iris.cnr.it/handle/20.500.14243/270337
Pubblicato in:
ANALYTICAL & BIOANALYTICAL CHEMISTRY (PRINT)