Multi-channel medical device for time domain functional near infrared spectroscopy based on wavelength space multiplexing
Articolo
Data di Pubblicazione:
2013
Abstract:
We have designed a compact dual wavelength (687 nm, 826 nm) multi-channel (16 sources, 8 detectors) medical device for muscle and brain imaging based on time domain functional near infrared spectroscopy. The system employs the wavelength space multiplexing approach to reduce wavelength cross-talk and increase signal-to-noise ratio. System performances have been tested on homogeneous and heterogeneous tissue phantoms following specifically designed protocols for photon migration instruments. Preliminary in vivo measurements have been performed to validate the instrument capability to monitor hemodynamic parameters changes in the arm muscle during arterial occlusion and in the adult head during a motor task experiment. © 2013 Optical Society of America.
Tipologia CRIS:
01.01 Articolo in rivista
Elenco autori:
Cubeddu, Rinaldo; Spinelli, LORENZO CLEMENTE
Link alla scheda completa:
Pubblicato in: