An insert for anelastic spectroscopy measurements from 80 K to 1100 K

Academic Article
Publication Date:
2009
abstract:
We describe an insert for measurements of the complex elastic moduli between liquid N2 temperature and over 1000 K. The system has been tested with free flexural vibrations of bar-like samples suspended on thermocouple wires with electrostatic excitation and capacitive detection of the vibration, providing excellent measurements with elastic energy loss coefficients between 3 × 10-6 and 0.1. The specially designed system for regulating the distance between the electrode and the sample would also allow optical detection of the vibration through optic fibers.
Iris type:
01.01 Articolo in rivista
Keywords:
Internal friction; stress relaxation; and mechanical resonances; Elasticity and anelasticity
List of contributors:
Latino, PAOLO MASSIMILIANO; Cordero, Francesco; Corvasce, Fabrizio
Authors of the University:
CORDERO FRANCESCO
CORVASCE FABRIZIO
LATINO PAOLO MASSIMILIANO
Handle:
https://iris.cnr.it/handle/20.500.14243/34911
Published in:
MEASUREMENT SCIENCE & TECHNOLOGY (PRINT)
Journal
  • Overview

Overview

URL

http://iopscience.iop.org/0957-0233/20/1/015702