Improving boston type 1 keratoprosthesis procedure: One-touch femtosecond-assisted preparation and centration of donor carrier tissue

Articolo
Data di Pubblicazione:
2014
Abstract:
Purpose: We describe a technique to avoid decentration of the visual axis of the Boston type 1 keratoprosthesis (Kpro), performing 2 concentric trephinations with femtosecond laser. Methods: Two concentric side cuts were performed in a donor cornea using the 150-kHz Intralase(TM) FS laser. Within the same applanation procedure, an 8.5-mm-diameter anterior side cut was performed, followed by a concentric 3-mm-diameter anterior side cut. Results: The technique was successfully replicated in 7 cases. Conclusions: Femtosecond laser-assisted double trephination results in a correctly prepared donor cornea, and in an inner side precisely matched with the prosthesis. At the end of the surgery, the Kpro was correctly centered. © 2013 Wichtig Editore.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Boston type 1; Femtosecond laser; Keratoprosthesis
Elenco autori:
Rossi, Francesca; Pini, Roberto
Autori di Ateneo:
ROSSI FRANCESCA
Link alla scheda completa:
https://iris.cnr.it/handle/20.500.14243/267274
Pubblicato in:
EUROPEAN JOURNAL OF OPHTHALMOLOGY
Journal
  • Dati Generali

Dati Generali

URL

http://www.scopus.com/inward/record.url?eid=2-s2.0-84892707386&partnerID=q2rCbXpz
  • Utilizzo dei cookie

Realizzato con VIVO | Designed by Cineca | 26.5.0.0 | Sorgente dati: PREPROD (Ribaltamento disabilitato)