Abstract:
The products of fusion reactions at JET are measured using different diagnostic techniques. One of the methods is based on measurements of gamma-rays, originating from reactions between fast ions and plasma impurities. During the forthcoming deuterium-tritium (DT) campaign a particular attention will be paid to 4.44 MeV gamma-rays emitted in the Be-9(alpha,n gamma)C-12 reaction. Gamma-ray detectors foreseen for measurements in DT campaigns have to be able to register spectra at high count rates, up to approximately 500 kHz. For the Gamma-ray Camera at JET a new setup will be based on scintillators with a short decay time, e.g., CeBr3, and a multi-pixel photon counter (MPPC). We present two methods of output signal shortening in modules based on MPPC. A short detector output signal is necessary in order to minimize the number of pile up events at high count rates. One method uses a passive RC circuit with a pole zero cancellation, whereas an active transimpedance amplifier is used in the other one. Due to the strong dependence of MPPC properties on temperature variation, a special device MTCD@NCBJ was designed and produced to stabilize the gain in MPPC-based scintillation detectors. We show that this device guarantees stable working conditions. (C) 2017 Elsevier B.V. All rights reserved.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Gamma-ray spectrometry; Scintillators; MPPC; Energy resolution; Decay time; Tokamak