Data di Pubblicazione:
2006
Abstract:
In this letter it has been proved that the vibrating resonance frequency of an atomic force microscope cantilever is strictly characterized by its thickness (alpha), while its width/thickness ratio (beta) appears to be a less sensitive parameter that can be approximated to a constant. We therefore propose a data analysis method that, by accounting for a constant beta, allows for the determination of the value of alpha and consequently to calculate eta. This method of monitoring viscosity has the advantage of requiring short measurement times on very small sample volumes, thereby avoiding laborious, time-consuming cantilever calibration.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Viscosity measurement; Atomic force microscopy; Cantilevers
Elenco autori:
Tiribilli, Bruno; Vassalli, Massimo
Link alla scheda completa:
Pubblicato in: