Problems of determining true equilibrium pressures of hydrogen getters over a wide range of getter temperature and hydrogen sorption

Articolo
Data di Pubblicazione:
1999
Abstract:
We have assessed a conventionally configured apparatus to measure hydrogen equilibrium pressures of getter alloys over the temperature range of 25 to 400 °C and hydrogen concentration range of 0.02 to 1 H/A. An in situ approach has been proposed to correlate the measurements of different gauges using one of the precisely calibrated gauges as a standard reference. The pressure range of interest was 10-7-105 Pa and was covered using a Bayart-Alpert gauge, a spinning rotor gauge, and different capacitance gauges. The approach involves measuring the characteristic response of these instruments prior to determining hydrogen equilibrium pressures over the relevant range of wall and getter bed temperature. To evaluate the accuracy of the measurements, we performed an error propagation analysis. The use of a porous filter to avoid fines contamination and the adsorption/desorption kinetics of the walls of the system were both addressed.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
SPINNING ROTOR GAUGE; IONIZATION GAUGES; SOLUBILITY; STABILITY; ISOTOPES
Elenco autori:
DE ANGELI, Marco; Ghezzi, FRANCESCO MAURO
Autori di Ateneo:
DE ANGELI MARCO
GHEZZI FRANCESCO MAURO
Link alla scheda completa:
https://iris.cnr.it/handle/20.500.14243/145130
Pubblicato in:
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY. A. VACUUM, SURFACES, AND FILMS
Journal
  • Dati Generali

Dati Generali

URL

http://avspublications.org/jvsta/resource/1/jvtad6/v17/i6/p3452_s1
  • Utilizzo dei cookie

Realizzato con VIVO | Designed by Cineca | 26.5.0.0 | Sorgente dati: PREPROD (Ribaltamento disabilitato)