Signal Processing and Calibration Procedures for in-situ Diode-Laser Absorption Spectroscopy
Articolo
Data di Pubblicazione:
2004
Abstract:
Gas analyzers based on tunable diode-laser spectroscopy (TDLS) provide high sensitivity, fast response and highly specific in situ measure- ments of several atmospheric trace gases simultaneously. Under optimum conditions even a shot noise limited performance can be obtained. For field applications outside the laboratory practical limitations are important. At ambient mixing ratios below a few parts-per-billion spectrometers become more and more sensitive towards noise, interference, drift effects and background changes associated with low level signals. It is the purpose of this review to address some of the problems which are encountered at these low levels and to describe a signal processing strategy for trace gas monitoring and a concept for in situ system calibration applicable for tunable diode-laser spectroscopy. To meet the requirement of quality assurance for field measurements and monitoring applications, procedures to check the linearity according to International Standard Organization regulations are described and some measurements of calibration functions are presented and discussed.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Diode laser spectroscopy; Gas analysis; Signal processing
Elenco autori:
D'Amato, Francesco; DE ROSA, Maurizio; Mazzinghi, Piero