Cadmium-Zinc-Telluride photon detector for epithermal neutron spectroscopy-pulse height response characterisation
Articolo
Data di Pubblicazione:
2004
Abstract:
The Resonance Detector Spectrometer was recently revised for neutron spectroscopic studies in the eV energy region.
In this technique one makes use of a photon detector to record the gamma emission from analyser foils used as neutrongamma
converters. The pulse-height response of a Cadmium-Zinc-Telluride photon detector to neutron capture
emission from 238U and 197Au analyser foils was characterised in the neutron energy range 1-200 eV: The experiment
was performed on the VESUVIO spectrometer at the ISIS neutron-pulsed source. A biparametric data acquisition,
specifically developed for these measurements, allowed the simultaneous measurements of both the neutron time of
flight and g pulse-height spectra. Through the analysis of the g pulse-height spectra the main components of the signal
associated with resonant and non-resonant neutron absorption were identified. It was also shown that, in principle,
energy discrimination can be used to improve the signal to background ratio of the neutron time-of-flight measurement.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
Neutron spectroscopy; Epithermal neutron; Resonance detector; CZT
Elenco autori:
Tardocchi, Marco; Pietropaolo, Antonino