Design and Simulation of a Helmholtz Coil for Magnetic Resonance Imaging and Spectroscopy Experiments with a 3T MR Clinical Scanner
Academic Article
Publication Date:
2019
abstract:
Magnetic resonance imaging (MRI) and magnetic resonance spectroscopy (MRS) are non-invasive techniques for tissue characterization. MRI/MRS in small phantoms with a clinical magnetic resonance scanner requires the design and development of dedicated radiofrequency coils. This paper describes the simulation, design, and application of a H transmit/receive Helmholtz coil, suitable for MRI/MRS studies in small phantoms with a clinical 3T scanner. Coil inductance and resistance were analytically calculated by taking into account the conductors cross geometry while magnetic field and sample-induced resistance were calculated with magnetostatic approaches. Finally, the coil sensitivity was measured with the perturbing sphere method. Successively, a coil prototype was built and tested on the workbench and by acquisition of MRI and MRS data. Results show that such coil could provide a low cost and easy to build device for MRI/MRS experiments with a clinical scanner in small specimens.
Iris type:
01.01 Articolo in rivista
Keywords:
Magnetic Resonance coils
List of contributors:
Giovannetti, Giulio
Published in: