Portable laser-heating stand for synchrotron applications

Academic Article
Publication Date:
2009
abstract:
A compact, double-sided laser-heating system for diamond-cell synchrotron applications is described. The optical table, containing laser, spectrometer, and all optics for visual observation and measuring temperatures and pressures has an area of less than ½ m2 and weighs less than 20 kg. All components can be remotely controlled at micron levels with simple dc motors and pneumatic drives. The design allows quick alignment of the laser-heated hot spot with the x-ray beam and the spectrometer. The prealigned system can be set up at most synchrotron beamlines within about 1 h. We carried out measurements on a variety of materials above one megabar and up to over 4000 K at both the x-ray diffraction beamline ID 27 and the x-ray absorption beamline ID 24 at the European Synchrotron Facility. A new measurement of the melting temperature of iron by x-ray absorption spectroscopy is presented.
Iris type:
01.01 Articolo in rivista
List of contributors:
Trapananti, Angela
Handle:
https://iris.cnr.it/handle/20.500.14243/15086
Published in:
REVIEW OF SCIENTIFIC INSTRUMENTS
Journal
  • Overview

Overview

URL

http://rsi.aip.org/resource/1/rsinak/v80/i4/p045103_s1