Improvement of sensitivity in continuous wave near infra-red spectroscopy systems by using silicon photomultipliers

Academic Article
Publication Date:
2016
abstract:
We experimentally analyze the signal-to-noise ratio of continuous wave (CW) near infrared spectroscopy (NIRS) reflectance systems based on light emitting diodes and silicon photomultipliers for high performance low cost NIRS biomedical systems. We show that under suitable experimental conditions such systems exhibit a high SNR, which allows an SDS of 7 cm, to our knowledge the largest ever demonstrated in a CW-NIRs system. (C) 2016 Optical Society of America
Iris type:
01.01 Articolo in rivista
Keywords:
SiPM; NIRS
List of contributors:
Pagano, Roberto; Lombardo, SALVATORE ANTONINO; Libertino, Sebania
Authors of the University:
LIBERTINO SEBANIA
LOMBARDO SALVATORE ANTONINO
PAGANO ROBERTO
Handle:
https://iris.cnr.it/handle/20.500.14243/335947
Published in:
BIOMEDICAL OPTICS EXPRESS