Performance of the prototype LaBr3 spectrometer developed for the JET gamma-ray camera upgrade
Academic Article
Publication Date:
2016
abstract:
In this work, we describe the solution developed by the gamma ray camera upgrade enhancement project to improve the spectroscopic properties of the existing JET ?-ray camera. Aim of the project is to enable gamma-ray spectroscopy in JET deuterium-tritium plasmas. A dedicated pilot spectrometer based on a LaBr3 crystal coupled to a silicon photo-multiplier has been developed. A proper pole zero cancellation network able to shorten the output signal to a length of 120 ns has been implemented allowing for spectroscopy at MHz count rates. The system has been characterized in the laboratory and shows an energy resolution of 5.5% at E? = 0.662 MeV, which extrapolates favorably in the energy range of interest for gamma-ray emission from fast ions in fusion plasmas.
Iris type:
01.01 Articolo in rivista
Keywords:
Cameras; Gamma ray spectrometers; Photomultipliers; Spectrometers Deuterium-tritium plasmas; Energy resolutions; Gamma-ray cameras; Gamma-ray emission; Gamma-ray spectroscopy; Pole zero cancellation; Silicon photomultiplier; Spectroscopic property
List of contributors:
Murari, Andrea; Giacomelli, LUCA CARLO; Tardocchi, Marco
Published in: