Hydrogen sulfide measurements in air by passive/diffusive samplers and high-frequency analyzer: A critical comparison
Academic Article
Publication Date:
2016
abstract:
In this study, hydrogen sulfide (H2S) measurements in air carried out using (a) passive/diffusive samplers
(Radiello® traps) and (b) a high-frequency (60 s) real-time analyzer (Thermo® 450i) were compared in
order to evaluate advantages and limitations of the two techniques. Four different sites in urban environments
(Florence, Italy) and two volcanic areas characterized by intense degassing of H2S-rich fluids
(Campi Flegrei and Vulcano Island, Italy) were selected for such measurements. The concentrations of
H2S generally varied over 5 orders of magnitude (from 101e103 mg/m3), the H2S values measured with
the Radiello® traps (H2SR) being significantly higher than the average values measured by the Thermo®
450i during the trap exposure (H2STa), especially when H2S was <30 mg/m3. To test the reproducibility of
the Radiello® traps, 8 passive/diffusive samplers were contemporaneously deployed within an 0.2 m2
area in an H2S-contaminated site at Mt. Amiata (Tuscany, Italy), revealing that the precision of the H2SR
values was ±49%. This large uncertainty, whose cause was not recognizable, is to be added to that related
to the environmental conditions (wind speed and direction, humidity, temperature), which are known to
strongly affect passive measurements. The Thermo® 450i analyzer measurements highlighted the
occurrence of short-term temporal variations of the H2S concentrations, with peak values (up to 5732 mg/
m3) potentially harmful to the human health. The Radiello® traps were not able to detect such temporal
variability due to their large exposure time. The disagreement between the H2SR and H2STa values poses
severe concerns for the selection of an appropriate methodological approach aimed to provide an accurate
measurement of this highly toxic air pollutant in compliance with the WHO air quality guidelines.
Although passive samplers may offer the opportunity to carry out low-cost preliminary surveys, the use
of the high-frequency H2S analyzer is preferred when an accurate assessment of air quality is required. In
fact, the latter provides precise real-time measurements for a reliable estimation of the effective exposure
to hazardous H2S concentrations, giving insights into the mechanisms regulating the dispersion of
this air pollutant in relation to the meteorological parameters.