Prototipo di sensore a fibra ottica per il monitoraggio del pH del fluido interstiziale in pazienti in terapia intensiva
Altro Prodotto di Ricerca
Data di Pubblicazione:
2005
Abstract:
An optical fibre sensor for measuring the pH in interstitial fluid is described. Microdialysis is the approach followed for extracting the sample from the subcutaneous adipose tissue. The interstitial fluid drawn flows through a microfluidic circuit formed by a microdialysis catheter in series with a glass pH sensing capillary. The pH indicator (phenol red) is covalently immobilised on the internal wall of the glass capillary. An optoelectronic unit that makes use of LED and photodetectors is connected to the sensing capillary by means of optical fibres. Optical fibres are used to connect the interrogating unit to the sensing capillary. A resolution of 0.03 pH units and an accuracy of 0.07 pH units were obtained. Preliminary in vivo tests were carried out in pigs with alterated respiratory function.
Tipologia CRIS:
05.12 Altro
Keywords:
pH; optical sensor; interstitial fluid; microdialysis; intensive care
Elenco autori:
Giannetti, Ambra; Baldini, Francesco; Mencaglia, ANDREA AZELIO