A point of care (POC) device for sepsis

Abstract
Publication Date:
2012
abstract:
In this work we describe the performances of a fluorescence-based optical device developed for sepsis analysis capable to measure the concentration of PCT, CRP and neopterin in serum. The heart of the system is a two-piece polymethylmetacrylate (PMMA) chip, with microchannels through which the analysed sample flows. The fluorescence, which comes from the sensing layer where the specific biologic interaction takes place, is guided by PMMA waveguides which run along the microchannels to an optical fibre connected to a Hamamatsu optical spectrum analyser.
Iris type:
04.02 Abstract in Atti di convegno
List of contributors:
Domenici, Claudio; Giannetti, Ambra; Trono, Cosimo; Puntoni, Mariarita; Baldini, Francesco
Authors of the University:
BALDINI FRANCESCO
GIANNETTI AMBRA
TRONO COSIMO
Handle:
https://iris.cnr.it/handle/20.500.14243/319441