Integrated Mach-Zehnder micro-interferometer for gas trace remote sensing
Contributo in Atti di convegno
Data di Pubblicazione:
2006
Abstract:
The realisation process and the preliminary tests on the performances of an integrated Mach-Zehnder Interferometer
on LiNbO3 (Lithium Niobate) substrate is presented. The microsystem has been obtained by using medium mass Ion
Implantation on X-cut Lithium Niobate crystals. The interferometer is formed by integrated optical channel waveguides;
the phase shift between the two optical paths has been obtained, without moving parts. by applying a suitable electric
field. The whole device is 60mm long, has a O.5xlmm2 cross section and weights a few grams. The power consumption
is in the milliwatt range. In the present work results obtained in the spectral window (O.4jim-l .Iim) will be presented.
The performance of the device, evaluated on standard radiation sources, demonstrates that a spectral resolution better
than O.3nm can be obtained on 400nm spectral windows. The Micro-interferometer has been tested in laboratory with a
calibrated cell containing NO2 gas and has demonstrated sensitivity in the ppb range if suitable optical paths are used.
Its reduced dimensions and weights make these micro-systems ideal for a wide range of applications, spanning from
Space Technology, Earth observation for Environmentmonitoring, to Safety and Security applications.
Tipologia CRIS:
04.01 Contributo in Atti di convegno
Elenco autori:
Cerutti, Annalisa; Pennestrì, Giuseppe; Chiarini, Marco; Bentini, GIAN GIUSEPPE; Bianconi, Marco
Link alla scheda completa:
Titolo del libro:
Remote Sensing and Infrared Devices and Systems
Pubblicato in: