Performance characteristics of a low volume spray chamber with a micro-flow nebulizer for ICP-MS.
Articolo
Data di Pubblicazione:
2004
Abstract:
Inspired by the original work of Todol?ยด et al. in designing a torch integrated sample introduction system
(TISIS) for ICP-AES, the authors have modified this design for use with ICP-MS. A microflow concentric
nebuliser has been coupled with a single pass low volume spray chamber that can be clipped directly onto the
base of a standard ICP-MS torch. The memory effects are very low: a 10 mg l21 tune solution containing Co,
Ce and Tl was found to have washed out within 4 s. Matrix effects were reduced significantly across the mass
range for a 10-fold diluted sea-water sample, with the signal being suppressed by only 10-20% compared with
the response in ultra-pure water; detection limits were found to be higher than with the normal set up, with the
instrument tuned to ''robust'' conditions. Finally, it was shown that the low dead volume and low flow rate
meant that transient signals could be acquired easily, by flow injecting 100 ml of a standard and obtaining a flat
topped peak with a 50 s signal plateau.