Calibration issues and pre-processing chain of the TASI-600 airborne LWIR hyperspectral scanner

Academic Article
Publication Date:
2017
abstract:
The paper presents some advanced techniques to improve the calibration and pre-processing of the TASI-600 thermal hyperspectral sensor acquired by the Italian National Research Council - Institute of Methodologies for Environmental Analysis (CNR-IMAA). TASI-600 sensor has 32 spectral bands in the 8.0-11.5 ?m spectral range, with a swath of 600 pixels and an IFOV of 2 mrad. The sensor is provided with a custom Stirling Cycle cooled MCT detector (HgCdTe) specifically designed for the TASI-600 and the system includes licensed pre-processing software to radiometrically correct the data. Nevertheless, after the TASI became operative, there was evidence of some artifacts induced in the calibrated images by the preprocessing procedures. Among them, blinking pixels, calibration artifacts and sporadic striping noise required the development of dedicated correction procedures. ? 2013 IEEE.
Iris type:
01.01 Articolo in rivista
Keywords:
Blinking pixels; Hyperspectral thermal imager; Radiometric calibration
List of contributors:
Palombo, Angelo; Pascucci, Simone; Santini, Federico
Authors of the University:
PALOMBO ANGELO
PASCUCCI SIMONE
SANTINI FEDERICO
Handle:
https://iris.cnr.it/handle/20.500.14243/391091
Published in:
WORKSHOP ON HYPERSPECTRAL IMAGE AND SIGNAL PROCESSING, EVOLUTION IN REMOTE SENSING
Series
  • Overview

Overview

URL

https://www.scopus.com/inward/record.uri?eid=2-s2.0-85038592329&doi=10.1109%2fWHISPERS.2013.8080711&partnerID=40&md5=96b2ce3bce9c6e06598817aa0a8ae8a8
  • Use of cookies

Powered by VIVO | Designed by Cineca | 26.5.0.0 | Sorgente dati: PREPROD (Ribaltamento disabilitato)