Comment on "a multivariate model based on dominant factor for laser-induced breakdown spectroscopy measurements" by Zhe Wang, Jie Feng, Lizhi Li, Weidou Ni and Zheng Li, J. Anal. At. Spectrom., 2011, DOI: 10.1039/c1ja10041f
Academic Article
Publication Date:
2011
abstract:
In a recent paper, Wang et al. (J. Anal. At. Spectrom., 2011, DOI: 10.1039/c1ja10041f) presented a combination of traditional PLS method and traditional univariate model that should improve the quality of quantitative elemental analysis in LIBS experiments. In this comment, we demonstrate that the proposed method is, in fact, basically flawed and its predictions are worse than the ones obtained using a standard univariate approach based on a calibration curve.