Publication Date:
2012
abstract:
The performance of a single crystal diamond (4.7x4.7x0.5mm(3) active volume) detector was tested in the ISIS pulsed neutron beam using biparametric (time of flight and pulse height) data acquisition. Three characteristic regions in the biparametric spectra are observed: i) low pulse height events with very short time of flight induced by gamma-rays; ii) low pulse height events at longer flight times (i.e. neutron energies E-n > 3.5-6 MeV), possibly due to neutron elastic scattering off C-12; iii) events with large pulse height and flight times corresponding to E-n > 6 MeV mainly due to inelastic reactions such as C-12(n,alpha)Be-9 and C-12(n,n')3 alpha. The potential use of this detector is discussed in relation to the ChipIr neutron beam line for fast neutron irradiation of electronic components at the ISIS spallation source.
Iris type:
01.01 Articolo in rivista
Keywords:
Instrumentation for neutron sources; Neutron detectors (cold; thermal; fast neutrons); Neutron sources; Solid state detectors
List of contributors:
Gorini, Giuseppe; Tardocchi, Marco
Published in: