International comparison of free-field hydrophone calibrations in the frequency range 10 kHz to 315 kHz

Academic Article
Publication Date:
1999
abstract:
This paper describes a European comparison of free-field hydrophone calibrations in the frequency range 10 kHz to 315 kHz. A total of twelve participants from seven European countries took part by calibrating three reference hydrophones, with project coordination provided by the National Physical Laboratory (NPL), UK. The agreement between the results was generally encouraging, with a majority of the results lying within 1 dB of the grand mean. However, some large variations were observed, and the uncertainties in the calibrations were typically underestimated by the participants, with the maximum differences from the grand means almost invariably exceeding the quoted overall uncertainties. A discussion is given of possible reasons for the disagreement, and of useful further work identified as a result of this comparison.
Iris type:
01.01 Articolo in rivista
List of contributors:
Buogo, Silvano
Authors of the University:
BUOGO SILVANO
Handle:
https://iris.cnr.it/handle/20.500.14243/171507
Published in:
METROLOGIA (ONLINE)
Journal
  • Use of cookies

Powered by VIVO | Designed by Cineca | 26.5.0.0 | Sorgente dati: PREPROD (Ribaltamento disabilitato)