Publication Date:
2020
abstract:
This paper reports a system for the determination of positions and velocities of particles flowing in a rectangular-section microfluidic impedance cytometer. Vertical positions are measured electrically using coplanar electrodes integrated in the microchannel, while transverse positions are measured optically using a high-speed camera mounted on an inverted microscope. The system operates at high-throughput and could be used as a quantitative tool in inertial microfluidics and separation studies. ? 17CBMS-0001.
Iris type:
04.01 Contributo in Atti di convegno
Keywords:
Coplanar electrodes; Impedance cytometry; Inertial focusing; Microfluidics; Particle tracking
List of contributors: