A hybrid mock circulatory system: Testing a prototype under physiologic and pathological conditions

Academic Article
Publication Date:
2002
abstract:
This concept is applied here to represent left ventricular function. The resulting hybrid model is inserted into the existing closed loop model of circulation. The hybrid model reproduces ventricular function by a variable elastance numerical model. Its interaction with the hydraulic sections is governed by measuring left atrial and systemic arterial pressures and computing the left ventricular output flow by the resolution of the corresponding equations. This signal is used to control a flow generator reproduced by a gear pump driven by a DC motor.
Iris type:
01.01 Articolo in rivista
List of contributors:
DE LAZZARI, Claudio; Ferrari, Gianfranco