Precision impedance measurements by the three-voltage method with a novel high-stability multiphase DDS generator
Articolo
Data di Pubblicazione:
2003
Abstract:
The system for the intercomparison of unlike impedances of the Istituto Elettrotecnico Nazionale Galileo Ferraris, based on the three-voltmeter method and four-terminal pair definition of the standards, has now been provided with a new direct-digital synthesis generator having multiple outputs, isolation among the channels, and high amplitude and frequency stability. Experimental results obtained using the new generator are shown and compared with those achieved with the previous system configuration.
Tipologia CRIS:
01.01 Articolo in rivista
Keywords:
AC generators; electrical variables measurement; impedance measurement; standards; SINEWAVE GENERATOR; 3-VOLTMETER METHOD; SYSTEM
Elenco autori:
Galzerano, Gianluca
Link alla scheda completa:
Pubblicato in: