Publication Date:
2017
abstract:
In the last few years, mathematical models of lung
ventilation have often been used to support the Anesthesiologists
and Resuscitators choices in the mechanical ventilator parameters
setting. In this context, a real-time control strategy is
doubtless crucial to avoid the occurrence of induced pressurederived
trauma.
In the present work, we develop a first version of a simple but
realistic physiological lung ventilation mathematical model. The
patient-ventilator complex is taken into account by modeling
the pressure wave provided by the mechanical lung ventilator
as an external (control) input. With the aim of reaching
the correct amplitude for the pressure wave at the mouth
provided by the mechanical ventilator, hence limiting the risk
of Acute Respiratory Distress Syndrome (ARDS), we consider
two different scenarios: the patient who needs to be ventilated
as a consequence of having an insufficient respiratory drift
(assisted ventilation) and the patient without a spontaneous
breathing (controlled ventilation). An output-feedback control
law is proposed, based on the flow measurements provided by
the ventilator, and not exploiting the full knowledge of the model
equations and parameters.
The considered approach looks promising, since a preliminary
in-silico validation of the resulting patient-ventilator
system shows that the target value of the tidal volume is readily
tracked in both scenarios, without dangerous oscillations and
with limited control effort.
Iris type:
04.01 Contributo in Atti di convegno
Keywords:
healthcare and medical systems; modeling; control applications
List of contributors: