Skip to Main Content (Press Enter)

Logo CNR
  • ×
  • Home
  • People
  • Outputs
  • Organizations
  • Expertise & Skills

UNI-FIND
Logo CNR

|

UNI-FIND

cnr.it
  • ×
  • Home
  • People
  • Outputs
  • Organizations
  • Expertise & Skills
  1. Outputs

Equivalent Circuit of the Neuro-Electronic Junction for Signal Recordings from Planar and Engulfed Micro-Nano-Electrodes

Academic Article
Publication Date:
2018
abstract:
In the latest years, several attempts to develop extracellular microtransducers to record electrophysiological activity of excitable cells have been done. In particular, many efforts have been oriented to increase the coupling conditions, and, thus, improving the quality of the recorded signal. Gold mushroom-shaped microelectrodes (GM?E) are an example of nano-devices to achieve those requirements. In this study, we developed an equivalent electrical circuit of the neuron-microelectrode system interface to simulate signal recordings from both planar and engulfed micro-nano-electrodes. To this purpose, models of the neuron, planar, gold planar microelectrode, and GM?E, neuro-electronic junction (microelectrode-electrolyte interface, cleft effect, and protein-glycocalyx electric double layer) are presented. Then, neuronal electrical activity is simulated by Hspice software, and analyzed as a function of the most sensitive biophysical models parameters, such as the neuron-microelectrode cleft width, spreading and seal resistances, ion-channel densities, double-layer properties, and microelectrode geometries. Results are referenced to the experimentally recorded electrophysiological neuronal signals reported in the literature.
Iris type:
01.01 Articolo in rivista
Keywords:
HSPICE simulation; neuro-electronic junction; neuron; planar and mushroom-shaped microelectrodes
List of contributors:
Martinoia, Sergio
Handle:
https://iris.cnr.it/handle/20.500.14243/346955
Published in:
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS
Journal