Publication Date:
2020
abstract:
We describe the development and the first characterization of a compact trace-gas sensor based on cantilever photoacoustic spectroscopy (CEPAS). The sensor was characterized in order to find the optimal operating parameters (pressure, molecule absorption line and laser modulated frequency). N2O was selected as test molecule. A quality factor of 200 at 10 mbar of cell pressure were determined. Furthermore, the first test measurements showed a minimum detection level of hundreds of ppb with integration time of 30 ms.
Iris type:
04.01 Contributo in Atti di convegno
Keywords:
Photoacoustic spectroscopy; gas trace sensing; cantilever-enhanced PAS
List of contributors:
Canino, MARIA CONCETTA; LOPEZ GARCIA, Inaki; Roncaglia, Alberto; DE NATALE, Paolo; CANCIO PASTOR, Pablo; Borri, Simone; Bonafe', Filippo
Published in: